Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
Portable and mobile
communications equipment
should be separated from
Armband by no less than the
distances calculated/listed
below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P) 80 to
800 MHz
D=(7/EI)(Sqrt P)
800 MHz to 2.5 GHz
Where P is the max power
in watts and D is the
recommended separation
distance in meters.
Field strengths from fixed
transmitters, as determined
by an electromagnetic site
survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing
a transmitter.
Guidance and Manufacturer’s Declaration - Emissions
The Armband is intended for use in the electromagnetic environment specified below.
The customer or user of the Armband should ensure that it is used in such an environ-
ment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - Guidance
ESD IEC 61000-
4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood,
concrete, or ceramic tile. If
floors are synthetic, they r/h
should be at least 30%
EFT IEC 61000-
4-4
±2kV Mains
±1kV I/Os
N/A Mains power quality should be
that of a typical commercial or
hospital environment
±1kV Differential
±2kV Common
N/A
Voltage Dips/
Dropout IEC
61000-4-11
>95% Dip for
0.5 Cycles
60% Dip for 5
Cycles
30% Dip for 25
Cycles
>95% Dip for 5
Seconds
N/A Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Armband requires
continued operation during
power mains interruptions,
it is recommended that the
Armband be powered from an
uninterruptible power supply
or battery.
Power
Frequency
50/60H
Magnetic Field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
Guidance and Manufacturer’s Declaration - Immunity
The Armband is intended for use in the electromagnetic environment specified below.
The customer or user of the Armband should ensure that it is used in such an environ-
ment.
User Environment (continued)